Cleaning the instrument exterior : Periodically wipe the instrument with a barely-damp cloth. User either warm soapy water
or isopropyl alcohol to dampen the cloth.
•
Inspect sample probe daily
•
If sample probe shows sign of fouling, it must be cleaned by gently wiping with a lint-free cloth dampened with isopropyl alcohol or
cleaning solution for protein removal, such as Thermo Scientific Orion Cleaning Solution 900021.
•
Why?
A dirty sample probe may affect accuracy and repeatability.
•
Clean cooling well weekly
•
Use Swab Cleaner Kit dampened with isopropyl alcohol to clean cooling well.
•
Why?
Buildup in cooling well may affect accuracy and repeatability and may cause “Sample Did Not Freeze” errors.
•
Examine exhaust fans monthly
•
Ensure exhaust fans are unobstructed by dust or debris.
•
Why?
Dirty exhaust fans will cause internal instrument temperature to rise, which may cause test errors or component failures.
•
Cleaning the Solenoid Semiannual
•
Instruments used daily should be cleaned monthly, while instruments that are used less frequently should be cleaned every three months.
•
Why?
A dirty solenoid can cause “Sample Did Not Freeze” errors and can affect instrument accuracy and repeatability.
Maintenance
Refer to User Guide for more information.
Содержание OsmoTECH PRO
Страница 1: ...Installation Training Guide OsmoTECH PRO Osmometer MP00082 Rev 0 PCN01026 ...
Страница 12: ... Major components Instrument Operation ...
Страница 13: ... Touchscreen Home Screen Instrument Operation ...
Страница 14: ... Touchscreen Legend Information Instrument Operation ...
Страница 16: ... Daily procedure Instrument Operation ...
Страница 38: ...Accesory 38 EPSON Dot matric printer 135022 Dot Matrix Printer Paper AN2TP5 ...