Advanced Bionics does not patch the Naída CI M90 sound processor. Firmware updates may be made
available from time to time and will require visiting your cochlear implant professional to reimage your
sound processor. Authenticity of firmware can be verified by your cochlear implant professional using
the Target CI fitting software.
Available Training
Users of the Naída CI M90 sound processor are trained by their cochlear implant professional at the
initial fitting session.
End of Cybersecurity Support
Advanced Bionics products are supported until they no longer have regulatory approval in the country of
use or until the product is obsolesced by Advanced Bionics.
GUIDANCE AND MANUFACTURER’S DECLARATION:
Per IEC 60601-1-2
Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions CISPR 11
Group 1
The Naída CI M90 sound processor uses RF
energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The Naída CI M90 sound processor is suitable
for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
Electromagnetic Immunity
The Naída CI M90 sound processor is intended for use in the electromagnetic environment specified
below. The customer or the user of the Naída CI M90 sound processor should assure that it is used in
such an environment.
Immunity
IEC 60601 Test Level
Compliance Level
a
Electromagnetic
Environment - Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact ± 8 kV air ± 6 kV contact ± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,