43
e
miSSioNS
t
eSt
C
omPliaNCe
e
leCtromagNetiC
e
NViroNmeNt
- g
uidaNCe
RF emissions
CISPR 11
Class B
The Naída CI M30 sound
processor is suitable for use in
all establishments, including
domestic establishments and
those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes. The Naída CI M30
is not intended for high EM
environments such as a MRI
scanner room or near HF
surgical electrocautery. Review
the Electromagnetic Immunity
section for determining safe
operating environments.
Harmonic
emissions IEC
61000-3-2
Not
applicable
Voltage
fluctuations/flicker
emissions IEC
61000-3-3
Not
applicable
Electromagnetic Immunity
The Naída CI M30 sound processor is intended for use
in the electromagnetic environment specified below. The
customer or the user of the Naída CI M30 sound processor
should assure that it is used in such an environment.