Adtech MVT 306L Скачать руководство пользователя страница 4

9.0  SPECIFICATIONS 

INPUT/OUTPUT 

INPUT SIGNALS 

a. 0.5 mv to 100 mv span (Z in greater than 10 megohm)

b. Zero Suppression:  up to 100% of the major range selected in

16 divisions of the coarse zero adjustment switch. 

c.  Span: from 0.5 mv to 100 mv full scale-switch selectable. The

coarse span switch adds 16 divisions to each major range.

d. Upscale/Downscale protection optional.

OUTPUT SIGNALS 

0-5 vdc, 1-5 vdc.

OUTPUT LOAD (AL)= 1 Ok Q min. 

PERFORMANCE 

a. 

Calibrated Accuracy: 

± 0.1 % 

b. 

Independent Linearity: 

± 0.01 % maximum,± 0.006% typical 

(14 bit digital linearity) 

c. 

Repeatability: 

± 0.005% maximum,± 0.002% typical 

d. 

Zero TC: 

±( input 

i�: 

(MV)  + 0.005 ) % of span max/

°

C

e.

Span TC: 

± 0.008% of span max/

°

C

f.

Load Effect: 

± 0.005% zero to full load 

g. 

Output Ripple: 

1 0 mv P/P maximum

h.

Response Time: 

11 O milliseconds (10 to 90% step response)

i. 

Bandwidth: 

(-3 db): 3.2 Hz

j.

Temperature Rang

0�2

° 

to 185

°

F (-31 

° 

to 85

°

C) operating 

. _,.j

to 200

°

F (-40

° 

to 93

°

C) storage 

k. 

Power Supply Effect: 

± 0.005% over operating range 

I. 

Isolation: 

(MVT 326 only) lnpuVOutput 600 vac

1000 vdc 

Note:  All accuracies are given as a percentage of span 

POWER 

a.  7 to 42 vdc-3.5 mA typical, 5 mA maximum .. 

MECHANICAL 

a. 

Electrical Classification: 

general purpose

b.

Connection: 

Screw, compression type, accepts up to 14 

AWG 

c. 

Controls: 

One 4 position jumper for major range

d. 

Mounting: 

e.  Weight: 

Two 16 position rotary switches for 

COARSE 

ZERO 

and SPAN control 

Two multiturn potentiometers for 

FINE 

ZERO and 

SPAN control 

DIN, Surface, Snap-Track, or NEMA 4 & 7 

Net Unit: 4 oz. (115 grams) 

Shipping: Nominal 7 oz.  (200 grams) 

OPTIONS 

Option Number 

H 13 through H 27 

14

Description 

Mounting 

Voltage/Current Inputs 

10. OUTLINE MOUNTING

"G" RAIL MOUNTING  � 

DIN EN 50 035 (32mm, 

1

5mm) 

1' 

-.---.---r--ir:===

3.

=

1

� 

3.55 

3-36 

3.65 

SLOT FORT  RA IL 

RELEASE 

"TRAIL  MOUNTING 

WIRE INSERTION 

SCREW CLAMP 

FROM TOP 

UPTO14AWG. 

DIN EN 50 022 (35mm, 

7.5mm) 

"G"  RAIL  MOUNTING 

DIN EN 50 035 (32mm, 

15mm) 

T RAIL  MOUNTING 

DIN EN 50022135mm. 15mml 

-4-

190-A-000083

Содержание MVT 306L

Страница 1: ...lable Installation area location must agree with thesuppliedinstrumentsincludingoperatingtemperature and ambient conditions Electrical Connections The wire used to connect the instrument to the control system 1 0 should be a twisted pair s and sized according to normal practice Shielded cableis not normallynecessary if used the shield must be grounded at the input negative of the ADTECH instrument...

Страница 2: ... is located on the input pcb furthest from the fine potentiometer controls The COARSE SPAN switch SW2 is located to the right of the Coarse Zero Switch Note Steps D G E are applicable when the unit is opened for re ranging to different Zero or Span E Apply an input corresponding to the zero input Adjust the fine ZERO control fully counterclockwise Adjust the COARSE ZERO switch to a position where ...

Страница 3: ...TABLE 2 Zero Elevation 0mV I J3 B 20mV I J3 A Note All input jumpers located on larger pcb 8 0 TABLES 8 1 PCB LAYOUT MVT 326L Isolated Millivolt 3 OUTPUT TABLE 2 Output Range 1 5 vdc I J6 A 0 5 vdc I J6 B Note J6 is on the larger pcb MVT 306L Non Isolated Millivolt 190 A 000083 ...

Страница 4: ...31 to 85 C operating _ jo to 200 F 40 to 93 C storage k Power Supply Effect 0 005 over operating range I Isolation MVT 326 only lnpuVOutput 600 vac 1000 vdc Note All accuracies are given as a percentage of span POWER a 7 to 42 vdc 3 5 mA typical 5 mA maximum MECHANICAL a Electrical Classification general purpose b Connection Screw compression type accepts up to 14 AWG c Controls One 4 position jum...