HSL-AI16AO2
Calibration
B.2 Calibrating the modules
To calibrate the modules:
1. Press the Calibration button from the HSL-AI16AO2 util-
ity. A dialog box appears.
2. Connect AI channel 1 to AGND. The AI channel index is
from 0 to 15. Take note of the index. After wiring, press
the AI Offset Calibration button.
3. Connect AI channel 0 to the calibrator, then, press the AI
Span Calibration button.
4. Connect AI channel 12 to AO channel 0, and AI channel
14 to AO channel 1, then press the AO Offset Calibration
button.
5. If the previous step is successful, press the AO Gain
Calibration button to finish the calibration.
If calibration is successful, the module is ready for use. If not,
check the wiring and calibrator, then repeat the calibration proce-
dures.