User manual soft air® wds · Version 09/2018
Once set, the pressure setting will be saved and will be retained until the pressure
setting is changed again. In the event of a power failure, the pressure setting is also
retained, and after the power is back on, the pump will automatically operate at the
last set pressure.
8 Alternating function
9 memory-function
In the case of the contraindications for the alternating pressure mentioned in the
indication table, the mattress pressure can be switched from alternate to static soft
storage by pressing button 4 (Static / Alternate) on the pump, during which the
display changes the operating mode from blue (alternate) to green (static). A comfort
setting is still possible in this mode. Since all cells support the static mode, a reduction
of 10-20 mbar from the set value is necessary.
To switch on the static function, press the button 4 , which is the middle blue button
on the right edge of the control panel. The display changes to the green LED (static).
To turn off the static function, press this button again. The display changes back to the
blue LED (Alternate). Please do not forget to increase the pressure by approx. 10 - 20
mbar, if you switch on the alternating pressure function again.