User Manual
soft air
®
wds
Art. No. 110 400-wds
Pressure ulcer prevention and treatment-system
Assembly and Operation instructions
Health. Security. Independence.
Страница 1: ...User Manual soft air wds Art No 110 400 wds Pressure ulcer prevention and treatment system Assembly and Operation instructions Health Security Independence...
Страница 2: ...ction Page 5 5 Explanation of indication and functions Page 5 6 Installation and setting up the mattress Page 6 7 To set soft air wds Page 7 8 Alternating function Page 9 9 memory function Page 9 10 C...
Страница 3: ...c position or alternating pressure The soft air wds large cell system has been developed for pressure ulcer prevention and treatment up to high risk Stage 3 With the writing of this user manual the pr...
Страница 4: ...ombustible gases or liquids may enter into the system EXPLOSION DANGER Warning Note The compressor is under power as soon as the power supply plug is put in Before opening the housing of the compresso...
Страница 5: ...utions are different The system soft air wds is a three chamber alternating pressure system with variable cycle times of 5 10 15 and 20 minutes and visual and acoustic alarm The acoustic alarm is acti...
Страница 6: ...kets from below up in the fixation slot The power unit can also simply be placed underneath the bed Please note that the power unit should not be covered in any way as thermal damage may occur Connect...
Страница 7: ...ressure is set in the middle part of the scale around 30 40 mbar Pressure can be lowered by repeated pressing on the button down to around 10 mbar By repeated pressing on the right button a maximum pr...
Страница 8: ...ck effect is to avoid High extended pressure in the sacral area Steps Weight 1 Step 1 led light ca 30 40 kg 2 Step 2 led lights ca 40 50 kg 3 Step 3 led lights ca 50 70 kg 4 Step 4 led lights ca 60 80...
Страница 9: ...static soft storage by pressing button 4 Static Alternate on the pump during which the display changes the operating mode from blue alternate to green static A comfort setting is still possible in thi...
Страница 10: ...irm ground e g bu using a special resuscitation board 11 Cleaning Service instructions 10 CPR function During use the mattress and tubes should be cleaned weekly using a damp soft cloth and mild deter...
Страница 11: ...fectiveness on a regular basis Disinfection with cleaning on locations is also possible when instructions for use and process time are properly kept ADL GmbH prescribes safety related revisions of all...
Страница 12: ...tubing Check CPR valve Ex change possible defect cell The patient sinks too deep into the mattress Air cell defect Tubing defect Pressure not correctly set Check cells and tubing connections Check pre...
Страница 13: ...6 1 1 5 Warranty 2 years 2 years Electr connections 230 V 7 W Protection class Protection class ll Fuse 230 V F500mA slow Cover Polyurethane washable to 95 C can be used in dryer to 50 C Application P...
Страница 14: ...on 09 2018 ADL GmbH An der Germania Brauerei 1 48159 M nster Germany Phone 49 251 149896 0 Fax 49 251 149896 99 ADL GmbH Production and servicer center Im Fl rchen 1 99334 Amt Wachsenburg Germany Phon...