CAUTION
- 1
H
CAUTION
Always use the supplied 3-pin cable for the power cable.
If any other power cable is used, it may result in an electric
shock to the operator.
H
CAUTION
If this instrument is connected to other peripheral devices, be
sure that they are connected in the manner specified by the
manufacturer.
Only connect devices specified by the manufacturer.
Connection to non-specified devices may result in electric
shock to the operator by current of leakage.
Precautions regarding settings
H
CAUTION
When maintaining (cleaning, sterilizing, etc.) the stimulator, be
sure to turn the power off and unplug the power cable from the
power outlet first.
Failure to do so may result in an electric shock or malfunction.
H
CAUTION
The electronic stimulator has a high voltage output capacity of
up to
±
50 V.
Directly touching the output terminals or electrodes during
stimulation or when the power cord is connected during
inspection and the stimulator is operable may result in electric
shock.
Precautions regarding maintenance
This alerts the user to imminent death or serious injury arising
from misuse of the stimulator.
This alerts the user to possible death or serious injury arising
from misuse of the stimulator.
This alerts the user to possible injury or damage to property
arising from misuse of the stimulator.
H
DANGER
H
WARNING
H
CAUTION
Notes are also provided in this manual to alert the user to specific instructions or
recommendations.
H
CAUTION
Do not touch the connector pins. Discharge electrostatic
energy by touching a grounded metal part before connecting.
The instrument may be damaged by electrostatic energy.
This device is designed for experimental and research
applications and must not be used for therapeutic purposes.
WARNING
General