2
8
AD
INSTRUMENTS
short as possible.
•
The 2 terminals of grounding wire must make sure for well electric conducting,
and process for antirust.
•
It is prohibited that users use other devices as part of grounding wire
’
s electric
circuit
The section of the conjunction between grounding wire and device
’
s frame should be
equal or greater than 25mm2
2.2.4 Frame Grounding
All the machine frames should connect to protective copper strip. The grounding wire
should be as short as possible and avoid circling. The section of the conjunction
between grounding wire and grounding strip should be equal or greater than 25mm2.
2.2.5 Device Grounding
Connecting the device
’
s grounding rod to frame
’
s grounding strip with copper
wire.
2.3 Wire
’
s Connection
The grounding wire connective screw is located at the right of rear panel and
the power switch, fuse, power supply outlet is just beside ,whose order goes like
this, power switch is on the left ,power supply outlet is on the right and the fuse
is just between them.
•
Connecting Power Cord
User can insert one end into power supply outlet, while insert the other end to
AC
power.
•
Connecting Grounding Wire
When the device solely connects to protective ground, it should adopt
Содержание AD9330
Страница 1: ...AD INSTRUMENTS 1 Manual de Usuario AD Instruments Manual de usuario Encoder AD9330 ...
Страница 12: ...1 2 AD INSTRUMENTS ...
Страница 15: ...1 5 AD INSTRUMENTS ...
Страница 19: ...1 9 AD INSTRUMENTS Manual de Usuario AD Instruments AD9330 Encoder ...