EN
TEMP IP68 LoRaWAN 863-870 - Guide utilisateur / User guide version V3.0.2
56 of 59
FR
Once the cable has been removed, insert the
gland plug provided in the package to maintain
the watertightness of the product
3
Using a suitable screwdriver,
unscrew the terminal block to
detach the wires
Unscrew the cable
gland to loosen
the cable
4
5V + -
1 2 3 4
5V + -
1 2 3 4
Remove the
screws
1
2
1
Open the product by
pulling on the cable
gland
WARNING : do not
pull on the wires
2
2
1
5V + -
1 2 3 4
5. WIRING
5.1. Disconnect a probe
To disconnect a probe, please follow these instructions:
CAUTION: To maintain the IP 68 it is important to screw the cable glands fully and screw the screws with a PZ.1 head
and a tightening torque of 0.9 N.m.
As a reminder the probe 1 is on the rounded side of the housing, the probe 2 is on the flat side of the housing.