Tare Function
Use the
TARE key
when the patient is weighed with any other equipment (such as carrying equipment,
belts, suspension yoke etc.). Weight the equipment first. After the weight is indicated press
Tare
key to
adjust the scale back to
zero
. The weight of the stored value will be deducted for all future weighing
procedures unless the tare function is used another time.
The weight displayed in the patient’s net weight.
GROSS/NET function
Using the
GROSS/NET key
you can jump between gross and net subject to a tare value has been
stored.
Zero setting
Pressing the
Zero key
you may
eliminate minor weighing irregularities which may occur after starting
the scale and the scale will be set to zero.
Printing function
In case a printer has been connected and the printing menu has been activated the<
> function is
sending the weight result to the printer. In case no printer has been connected the function is
interrrupted after 10 seconds.
„Lifeweight“ - function
The
f- key
is manually activated and deactivated dependant from the requirement to weigh extremely
restless patients and getting a reliable result. In case the Lifeweight function is activated the scale
identifies all major movements caused by the patient during weighing in either direction and is
automatically cutting these irregularities off. Different automatical measurements done during weighing
the final weight of the patient displayed is most accurate.
Display symbols
NET
Scale is under tare
~
(Motion)
Scale is in motion
S
(above limit) The limit has been exceeded
T
(below limit) Off limit
(Battery)
Batteries low – insert fresh ones