Changing automatic switch-off time 

In order to change the automatic switch-off time, it is necessary to open a secure menu.  

 

Open the menu only when you really have to. Accidental or 
intentional changes to the parameters can have a significant impact 
on the functionality of the scale. 

To enter the secure menu, press and hold down the 

 key until “CAL” is displayed as you 

switch on the scale and after it has switched on. 

The parameter menu has 12 functions. The display shows the names of functions (“CAL”, 
“rES”, “CAP”, “Cnt”, “GrA”, “CoM”, “bAUd”, “oFF”, “SPEEd”, “MtArE”, “rESt” and “bACk”). You 

can navigate through these functions by repeatedly pressing the 

 key. To select a function, 

press the 

 key. You can navigate through the values of this function by repeatedly 

pressing the 

 key. To save your selection, press the 

 key. 

The menu can be exited by selecting the “bACk” parameter and pressing the 

 key. 

Caution: On calibrated scales, various functions such as “CAL” and “GrA” are locked. A 

change leads to the loss of the calibration and is only reserved for qualified personnel 
authorised by the manufacturer. 

 

Indicator 

Description 

oFF 

Setting the automatic switch-off time. The automatic switch-off time can be 
set to 3, 5, 15 or 30 minutes (“oF 3”, etc.) or the automatic switch-off function 

can be disabled (“oF 0”). Press the 

 key to retrieve the currently set 

value. Press the 

 key to set the desired value. Press the 

 key to 

confirm the displayed value.

 

10. Care and maintenance 

Cleaning 

Disconnect the mains plug before cleaning the scale. Use only a damp cloth or an ordinary 
disinfectant for cleaning. Do not use aggressive liquid cleaning agents, abrasive or acidic 
detergents or white spirit.  

Make sure that no liquid cleaning agent or water penetrates the scale and always follow the 
manufacturer's instructions for use.  

 

 

Содержание GW400 Series

Страница 1: ...OPERATING MANUAL Electronic compact scales Compact scale GW400 series GW400 210408 Rev004 UM en...

Страница 2: ...6 Key names and functions 6 Display symbols 7 Battery status indicator 7 8 Using the scale 7 Preliminary note 7 Switching the scale on and off 7 Zeroing 8 Taring 8 9 Backlight and parameter settings...

Страница 3: ...ne When reselling the scale the operating manual must also be supplied Target group This operating manual is intended for personnel who are responsible for setting up operating care and maintenance of...

Страница 4: ...4 Intended use Your ADE scale is a quality product and was manufactured for use as a non automatic scale The scale is intended for determining the weight of goods and for a variety of other weighing...

Страница 5: ...vibrating table as this will cause distortion of measurement results 4 Connect one end of the power supply to the power connector on the bottom of the scale and the other end to the local power supply...

Страница 6: ...y connector 3 Spirit level 8 On Off switch 4 Keypad 9 Battery compartment cover 5 Battery indicator Key names and functions Symbol Description Function ZERO Zeroing key If this key is pressed the scal...

Страница 7: ...to prevent irreparable damage to the load cell Switching the scale on and off Switching on the scale Remove all weights from the weighing surface Switch on the scale The On Off switch is located belo...

Страница 8: ...ce the tare weight will be displayed as a negative value Pressing the key after unloading the weighing surface clears the tare weight If the display is not stable the arrow will not light up on the st...

Страница 9: ...n press the key The menu can be exited by selecting the bACk parameter and pressing the key Caution On calibrated scales various functions such as CAL and GrA are locked A change leads to the loss of...

Страница 10: ...cannot be eliminated or other error messages occur please contact your service partner Faults Fault Cause Measure The display shows nothing The scale has switched off automatically Switch on the scale...

Страница 11: ...f the scale to avoid damage Disposal Waste electrical equipment does not belong in household waste Devices with this marking may not be disposed of as residual waste but must be recycled Remove if pos...

Страница 12: ...authorised personnel can perform recalibration On the CE mark you can find the year of initial calibration next to the notified body 0122 Recalibration must be carried out when one or more security m...

Страница 13: ...LED backlight display backlight can be disabled Casing Stable plastic housing Keypad Easy care membrane keypad Functions Weighing taring Automatic zeroing Overload indicator Battery level indicator a...

Страница 14: ...NAWI directive 2014 31 EU EMC Directive 2014 30 EU Low voltage directive 2014 35 EU RoHS Directive 2011 65 EC in their respectively applicable versions This declaration will lose its validity if any...

Страница 15: ...GW400 210408 Rev004 UM en 15...

Страница 16: ...16 GW400 210408 Rev004 UM en Distributor...

Отзывы: