14
backlight shutdown t time, 1/5/10 minutes are optional, or ON (the backlight is always on).
8.5 Measurement Filter Parameter Setting
In the filter adjustment menu if the measurement is voltage, current or loop transmitter
current, the filter coefficients will be displayed directly. If there are other items, you can press key to switch between
measurement items.
Press key or key to select first order filtering coefficients, 2…10 are optional (the larger the value, the more
obvious the filtering effect) or OFF (no filter).
8.6 Automatic Ramp/Step Output Setting
8.6.1 Ramp/Step switch
In the
menu press key or key to switch between
automatic ramp/step setup
automatic ramp output
setting
automatic step output setting
and
.
8.6.2 Ramp Direction Setting
In the
menu press to set
.
automatic ramp/step setting
automatic ramp/step direction
This icon means the direction is 0%→100%→0% or 100% →0%→100%, and means the 0%→100%
or 100%→0%.
8.6.3 Ramp/Step Time Setting
In automatic ramp/step setting menu press key to enter the time setting. Press key to move the cursor
and press key to increase or decrease by 1 second.By pressing the key again will confirm the entry or
Mode
Auto
Mode
100%
Math
Hold
Auto
Hold