Accuracy
0.06%,0.1%,0.16%,0.25%,0.4%,
0.6%,1%,1.6%,2.5%,4%,custom
Accuracy of DUT, the output is customizable
Transfer function
Linear / square / square root
Transfer function of the DUT signal
Owner
Numbers, letters, symbols
Owner of DUT
Location
Numbers, letters, symbols
Location of DUT
Note
Numbers, letters, symbols
Additional information
3. Task setting
Refer to the task setting in paragraph 6.1.1.3.
4. Start the task
For part of the content, refer to Start the task in paragraph 6.1.1.4.
The calibrator will automatically record the measured signal value at the moment, the current display of DUT
needs to be input manually.
5. Finish the task
Refer to Finish the task in paragraph 6.1.1.5.
6.4 Calibrate loop devices
6.4.1 Loop indicators
1. Connect to DUT
Connect the DUT input with the electrical output module of the calibrator, as shown in the Figure 14(for current
output) or 13(for voltage output).
2. Create the task