8250A Optical Power Meter Operation Manual
3.2 Menu Description
3-3
3.2
Menu Description
Refer to 3.1, “Menu Map.”
Smoothing count
Sets the number of times that the smoothing calculation is per-
formed.
The setting range is from 0 to 100. (If 0 or 1 is set, the calculation
is set to OFF.)
CF calculation ON/OFF
Sets the CF calculation to ON or OFF.
CF
Sets the correction factor by which the measured value is divided.
Setting range: 0.001 to 999.999.
Refer to the formula in section 3.3.5, “Reading the Display.”
Calibration wavelength
Sets the calibration wavelength.
Selects the calibration wavelength which is used in the multi-
wavelength calibration option.
Figure 3-3 Example of Calibration wavelength Settings
Number of display digits
Sets the number of display digits of the measured value to 3, 4, or
5.
If the display digit is set to 3 or 4, an under-bar is displayed in
place of each unused digit.
Parameter save
Saves the current settings in the built-in flash memory.
The settings can be saved in the following 4 areas: USER-0,
USER-1, USER-2, and USER-3.
The flash memory is initialized when "dFLt" is selected.
For more information, refer to 3.4, “Backing Up and Initializing
Parameter load
Loads the settings which are saved in the built-in flash memory.
The settings, which are saved in the following 4 areas: USER-0,
USER-1, USER-2, and USER-3, can be loaded.
The current settings are initialized when "dFLt" is selected.
For more information, refer to 3.4, “Backing Up and Initializing
1
2
3
4
Wavelength number display
(Cursor blinks.)
Numerical value UP or DOWN
Wavelength display
Wavelength sensitivity coefficient
5
6
7