8250A Optical Power Meter Operation Manual
8.2 Optical Sensor Specifications
8-4
8.2
Optical Sensor Specifications
8.2.1
Sensor Specifications (Sold Separately)
823x series
Model
82311/82311B
(General-purpose)
82312/82312B
(For blue-violet light)
82313/82313B
(For high power light)
Wavelength range
390 to 1100 nm
390 to 450 nm
390 to 1100 nm
Power range
Display in dBm
-60 to +17 dBm
-50 to +20 dBm
-50 to +23 dBm
Display in W
1 nW to 50 mW
10 nW to 100 mW
10 nW to 200 mW
Beam spot
3 mm
or more
1 mm
or more
0.1 mm
or more
Light receiving element
Si photodiode
Light receiving area
Approx. 9.5 mm
9.5 mm
Approx. 10 mm
10
mm
Approx. 8.5 mm
Effective light receiving area *1
Approx. 8.5 mm
8.5 mm
Approx. 6 mm
Calibration wavelength *2
780 nm
405 nm
650 nm
Measuring accuracy (when 1 mW is input)
2.5% (at calibration wavelength)
(
3.5%) *3
(400 to 1000 nm)
3.5%
(390 to 450 nm)
3.5%
(400 to 1000 nm)
Wavelength sensitivity correction range
390 to 1100 nm
390 to 450 nm
390 to 1100 nm
Shape
Thin type
Separation from a sensor section *4
Impossible
Possible
Possible
Dimensions (width)
(height)
(thickness of
the light receiving section) mm
18
180
3.2
18
180
3.7
18
180
5
Model
82321/82321B
(General-purpose)
82322/82322B
(For blue-violet light)
82323/82323B
(For high power light)
Wavelength range
390 to 1100 nm
390 to 450 nm
390 to 1100 nm
Power range
Display in dBm
-60 to +17 dBm
-50 to +20 dBm
-50 to +23 dBm
Display in W
1 nW to 50 mW
10 nW to 100 mW
10 nW to 200 mW
Beam spot
3 mm
or more
1 mm
or more
0.1 mm
or more
Light receiving element
Si photodiode
Light receiving area
Approx. 8.5 mm
Approx. 8.5 mm
Approx. 8.5 mm
Effective light receiving area *1
Approx. 6.5 mm
Approx. 6.5 mm
Approx. 6 mm
Calibration wavelength *2
780 nm
405 nm
650 nm
Measuring accuracy (when 1 mW is input)
2.5% (at calibration wavelength)
(
3.5%) *3
(400 to 1000 nm)
3.5%
(390 to 450 nm)
3.5%
(400 to 1000 nm)
Wavelength sensitivity correction range
390 to 1100 nm
390 to 450 nm
390 to 1100 nm
Shape
Cylindrical type
Dimensions (diameter)
(height) mm
38
40
*1:
Relative sensitivity to the center is within the
10 % range.
*2:
Can be added by using options. (A calibration wavelength other than the standard specification can be
selected by using the calibration wavelength selection function.)
*3:
Only when 20 is specified
*4:
The warranty does not include cut cables and/or damaged or degraded elements caused by continuous con-
necting and disconnecting to the sensor section.