7351 Series Digital Multimeter Operation Manual
6.7.5 Status Register Structure
6-50
7. Operation event register
Allocation in the operation event register is shown in Table 6-9.
Conditions which clear the operation event register
• When the power is turned on.
• When the *CLS command is executed.
• When the :STATus:OPERation[:EVENt]? command is executed.
• When the OSR? command is executed.
Conditions which clear the operation event enable register
• When the power is turned on.
• When the :STATus:OPERation:ENABle 0 command is executed.
• When the OSE0 command is executed.
Table 6-9 Operation Event Register (OER)
bit
Name
Description
0
Not used
Always 0
1
Not used
Always 0
2
Not used
Always 0
3
Not used
Always 0
4
Not used
Always 0
5
Waiting for TRIG
ON: This bit is set to 1 when this instrument is in Trigger Layer.
6
Not used
Always 0
7
Not used
Always 0
8
Not used
Always 0
9
Idle
ON: This bit is sets to 1 when this instrument is in the Idle state.
10 Not used
Always 0
11 Not used
Always 0
12 Not used
Always 0
13 Not used
Always 0
14 Not used
Always 0
15 Not used
Always 0