4.2.6 MENU Key (Parameter Setting)
4-10
RANDOM MEM
Sets the source data (random memory) for random sweep.
Data Set..
Sets addresses and data in the random memory.
Addresses from 0 to 7999 are available.
Pressing the SEL key switches between the address and data set-
ting parameters.
Save/Clear
Saves or clears the random memory.
Save:
Saves the setting data in the random memory to the
non-volatile memory. The saved data is loaded at power
ON.
Clear:
Clears the setting data in the random memory.
However, the non-volatile memory data is not cleared.
COMPUTE
Set the calculation-related parameters.
Compare SW
Switches the comparator calculation ON or OFF.
On:
Executes the comparator calculation.
The calculation result is reflected on the
indicator,
the header of data output (Talker output) and the status
byte.
HI; High value < Measurement data
GO; Low value
Measurement data
High value
LO; Measurement data < Low value
Off:
Turns OFF the comparator calculation.
Scaling SW
Switches the scaling calculation ON or OFF.
Scaling calculation =
On:
Executes the scaling calculation.
The MATH indicator goes ON.
Off:
Turns OFF the scaling calculation.
Max/Min SW
Switches the MAX/MIN calculation ON or OFF.
On:
Executes the MAX/MIN calculation.
The MAX, MIN, AVE, and
indicators go ON.
Off:
Turns OFF the MAX/MIN calculation.
View Mx/Mn
Displays the MAX/MIN calculation data.
Sample
Number of measurement data
Maximum
Maximum measurement data value
Minimum
Minimum measurement data value
Average
Measurement data average value
Total
Total measurement data value
Constant A
(Measurement Value) - Constant B
Constant C