6240B DC Voltage Current Source/Monitor Operation Manual
6.4 USB
6-8
6.4
USB
6.4.1
Overview
The 6240B is equipped with USB (Universal Serial Bus) conforming to USB 2.0 standard.
USB allows function settings and reading of measurement data with respect to multiple instruments con-
nected to the bus from a PC, making it simple to configure an automated measurement system.
NOTE:
All operations using a PC or hub cannot be always guaranteed.
6.4.2
USB Specifications
•
Standard:
Complies with USB2.0 Full-Speed
•
Connectors:
USB type B (female)
•
ID:
USB ID, settable from 1 to 127
•
Remote/Local:
Available
•
Input commands:
Function setting and query with ASCII character string commands
•
Output format:
Measurement data and query response with ASCII character strings
•
Driver:
ADC measuring instrument USB driver
6.4.3
Setting Up USB
6.4.3.1
Connecting with PC
Connect the USB connector (type B) on the rear of the instrument to the PC USB connector with a cable.
Fully insert all connectors.
When connecting multiple instruments to a single PC, use a USB hub.