4601 I-V Meter Operation Manual
1.6 Operating Environment
1-11
1.6
Operating Environment
This section describes the required environmental and power supply conditions.
1.6.1
Environmental Conditions
The 4601 must be installed in an environment meeting the following conditions:
•
Operating environment:
Ambient temperature 0
C to +50
C
Relative humidity 85% or lower without condensation
•
Storage environment
Ambient temperature -25
C to +70
C
Relative humidity 85% or lower without condensation
•
Location not subject to corrosive gasses
•
Away from direct sunlight
•
Dust free
•
Vibration free
•
Noise free
The 4601 is designed with full consideration given to the noise contained in the AC power line.
Nevertheless, it is recommended that the 4601 be used in an environment with as little line noise as
possible.
If line noise is unavoidable, use a noise filter.
•
Positioning of the 4601
A cooling fan is located on the rear panel and vents are located on the side panels. Blocking the fan
or vents cause the internal temperature to rise, possibly causing faulty operation.
Leave at least 10 cm of free space between the rear panel and the wall. Also, do not position the 4601
with the rear panel facing down.
•
Mounting in a rack
Ensure that exhaust air from other devices is not directed at the vents on the side of the 4601.
To prevent the temperature in the rack from rising, install a heat sink fan.
Содержание 4601 I-V Meter
Страница 2: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 70: ...4601 I V Meter Operation Manual 3 4 1 Menu Structure 3 18...
Страница 80: ......
Страница 92: ......
Страница 129: ...4601 I V Meter Operation Manual 5 6 1 Synchronous Operation 5 37...
Страница 146: ......
Страница 182: ......
Страница 230: ......