4601 I-V Meter Operation Manual
9.8 Setting Time
9-9
9.8
Setting Time
Minimum step (repeat) time:
Voltage/current/reference current measurement in fixed source/
measurement ranges and with the integration time of 5
s and with th
minimum measurement/source delay time
Setting time
Mode
Minimum step time
Sweep
50
s
DC
5 ms
Setting time
Setting range
Minimum resolution
Setting accuracy
Source delay time
0.010 ms to 5999.8 ms
1
s
(0.1% + 10
s)
Period (cycle)
0.050 ms to 6000.0 ms
1
s
(0.1% + 10
s)
Measurement delay time
0.020 ms to 5999.8 ms
1
s
(0.1% + 10
s)
Hold time
0 ms to 6000.0 ms
100
s
(2% + 2 ms)
Trigger delay time
0 ms to 6000.0 ms
100
s
(0.1% + 100
s)