4601 I-V Meter Operation Manual
8.3 Calibration Points and Tolerance Ranges
8-6
8.3
Calibration Points and Tolerance Ranges
For calibration, use the measurement instruments satisfying the required accuracies described in Section
8.1.1, "Measuring Instruments Necessary for Calibration", meeting the tolerance ranges shown in the follow-
ing table.
Table 8-2 Calibration Tolerance Ranges
1. Voltage source/voltage measurement/current measurement terminal
Source calibration
Voltage measurement calibration
Voltage source calibration (VS)
Voltage measurement calibration
(VS/VM)
Range
Calibration point
Tolerance
specification [digit]
Range
Calibration point
Tolerance
specification [digit]
5 V
0.0000 V
1d
5 V
0.00000 V
2d
+5.0000 V
1d
+5.00000 V
2d
50 V
00.000 V
1d
50 V
00.0000 V
2d
+50.000 V
1d
+50.0000 V
2d
300 V
000.00 V
1d
300 V
000.000 V
2d
+300.00 V
1d
+300.000 V
2d