3-5
12704A Resistivity Chamber Operation Manual
3.1.4 Connecting with the 5450/5451
3.1.4
Connecting with the 5450/5451
See Figure 3-2.
(1)
Connect the INPUT terminal of the 5450/5451 to the INPUT terminal of the 12704A by
using the S.TRIAX-TRIAX cable (CC010004/CA01009) for the 5450 or the
TRIAX-TRIAX cable (A01009) for the 5451.
(2)
Connect the V SOURCE terminal of the 5450/5451 to the V SOURCE HI terminal of
the 12704A by using the output cable (CC010005).
(3)
When using the LID SIGNAL function of the 12704A, connect the INTERLOCK/LID
on the rear panel of the 5450/5451 to the LID SIGNAL connector of the 12704A by us-
ing the BNC-BNC cable (A01036). The INTERLOCK/LID function of the 5450/5451
operates differently depending on the setting. For more information on the function and
its setting method, refer to the 5450/5451 Operation Manual.
(4)
Fix the short bar of the 5450/5451 between the LO and GUARD terminals.
The 5450/5451 outputs applied voltage of maximum 1000 V. Perform measurement
with great care.
If a cable is disconnected with voltage being applied from the V SOURCE, voltage
generation will occur on the end of the cable. Be sure to set the voltage STANDBY
before disconnecting the cable.
12704A
CC010005
A140001
5450
:
CC010004/ CC A01009
5451
:
A01009
5450/5451
To INTERLOCK/LID terminal
on the rear panel
A01036
Figure 3-2 Connecting with the 5450/5451
WARNING
Содержание 12704A
Страница 2: ......
Страница 4: ......
Страница 5: ...1 1 12704A Resistivity Chamber Operation Manual 1 Introduction Chapter 1 Introduction ...
Страница 9: ...2 1 12704A Resistivity Chamber Operation Manual 2 Part Descriptions Chapter 2 Part Descriptions ...
Страница 14: ......
Страница 15: ...3 1 12704A Resistivity Chamber Operation Manual 3 Measurement Method Chapter 3 Measurement Method ...
Страница 23: ...4 1 12704A Resistivity Chamber Operation Manual 4 Specifications Chapter 4 Specifications ...
Страница 25: ...A 1 12704A Resistivity Chamber Operation Manual APPENDIX APPENDIX ...