Bent Arm Illuminator and Accessories
Troubleshooting Guide
Trouble Area
No light output
Accessories do not
attach securely or
operate properly
Instrument head will
not connect
Dim light output
Visible corrosion on
instrument
Possible Cause
Lamp burned out
Wrong lamp installed
Batteries are depleted
Accessories may have
reached end of useful
life
Thread/connection is
damaged
Lamp near end of
useful life
Batteries are depleted
Damage from excessive
moisture
Corrective Action
Replace the lamp with ADC part
#5211-4
Replace the lamp with ADC part
#5211-4
Replace with fresh “C” batteries
Obtain replacement accessories
Service required
Replace the lamp with ADC part
#5211-4
Replace with fresh “C” batteries
Service required