background image

A

Special

Thank

You...

Thank you for choosing an ADC blood pressure instrument. We’re proud of the care and quality that goes
into the manufacture of each and every sphygmomanometer that bears our name.

Only the finest materials are used to assure you of a timeless instrument designed for optimum perform-
ance.

You’ll quickly appreciate the results for you now own the finest sphygmomanometer that money can buy.

With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many
years of dependable service. Please read the following instructions and general information which will prove
helpful in allowing you to enjoy your ADC product.

Thank you for your patronage. It is indeed our pleasure to serve you.

                                                   

Sincerely,

                                                   

American Diagnostic Corp.

Device Description and Intended Use

ADC

®

aneroid sphygmomanometers are used by professional healthcare providers and individuals trained in auscultatory

blood pressure technique to determine systolic and diastolic blood pressure in humans and animals.

Warnings and Precautions

Warning

- A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately

could lead to patient injury, illness, or death.

Warning

- Do not allow a blood pressure cuff to remain on patient for more than 10 minutes when inflated above 10 mmHg.

This may cause patient distress, disturb blood circulation, and contribute to the injury of peripheral nerves.

Warning

- If luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently con-

nected to intravascular fluid systems, allowing air to be pumped into a blood vessel. Immediately consult a physician if this occurs.

Warning

For all blood pressure systems that can be wall mounted (Diagnostix

TM

952B), ensure that the unit is securely mounted

prior to use to avoid damage to the instrument and potential patient injury.

Warning 

-

Safety and effectiveness with neonate cuff sizes 1 through 5 is not established.

Caution 

Federal law restricts this device to sale by or on the order of a physician or licensed healthcare practitioner.

Warning

Do not twist connectors when joining the blood pressure cuff to tubing as this may cause the connector to stick and

make changing the blood pressure cuff difficult.

Warning:  

When using with an infant or child cuff, extra care must be taken to prevent over-inflation. With smaller cuffs (infant

or child) the cuff can inflate to over 300mmHg with just two full compressions of the bulb.To prevent discomfort or injury to the

patient and damage to the instrument, bulb should only be partially squeezed, so that each “stroke” inflates the cuff in 40mmHg

to 60mmHg increments until inflated to the desired level.

Warning

This Product May Contain A Chemical Known To The State of California To Cause Cancer, Birth Defects, Or Other Re-

productive Harm.

Отзывы: