background image

Maintenance:

 

Cleaning and disinfection procedures must be conducted by persons trained in medical device cleaning and disinfection. If 
using a germicidal or other cleaning agent, consult the manufacturer’s instructions for proper use and efficacy. 

 

CAUTION:

 Use only the cleaning or germicidal cleaner agent types listed below or damage to cuff may occur. 

CAUTION: 

Repeated reprocessing may cause degradation of device; follow inspection procedures to assure integrity of device. 

CAUTION: 

Do not aggressively scrub cuff as damage to cuff markings and/or cuff closure integrity may occur. 

 

Materials:

 

When cleaning or disinfecting, use sterile cloths, soft brush, soaking tray, and potable rinse water (softened preferred). 

 
For low-level disinfection, use one of the following: 

   •  Bleach-based germicidal cleaner suitable for use on healthcare equipment and capable of low-level disinfection; for 
       example, a 1:10 dilution of bleach (6500 ppm sodium hypochlorite) and detergent. 
   •  Commercial wipes with 0.5% bleach and water solution or 70% isopropyl alcohol.  

 
Reference EPA-registered disinfectants at: https://www.epa.gov/pesticide-registration/list-antimicrobial-products-registered-
epa-sterilizers. 

 

Basic Cleaning:

 

Note: 

Do not allow liquid to enter tubing while cleaning. 

   1. Use a soft brush to remove visible contamination. 
   2. Thoroughly rinse with water.  
   3. Damp dry with a clean cloth and allow to air dry further if needed. 

 

Low-Level Disinfecting:

 

Note: 

Do not allow liquid to enter tubing while disinfecting. 

   1. Use a soft brush to remove visible contamination. 
   2. Rinse with water.  
   3. Thoroughly saturate (spray or immerse) all surfaces of the cuff and accessories with selected cleaning agent.  
   4. Soft brush all surfaces. Allow a 5-minute wet contact time, or longer, if directed by the germicidal cleaner manufacturer. 
       Do not exceed 10 minutes of wet contact time. 
   5. Thoroughly rinse with water. 
   6. Damp dry with a clean cloth and allow to air dry further if needed. 

 
After cleaning or disinfecting, inspect cuff for deterioration and adequate closure integrity, and inflate to assess for leaks. Do 
not use the cuff if any abnormalities are found. 
 

Environmental Conditions:

 

 

Storage temperature:

          -4°F to 149°F (-20°C to 65°C)   

Operating temperature:

          32°F to 122°F (0°C to 50°C) 

Storage relative humidity:

  15% to 95% non-condensing      

Operating relative humidity:

  15% to 90% non-condensing

 

 

              ADC 
              55 Commerce Drive 
              Hauppauge, NY 11788 
              U.S.A.

Inspected & Packaged in the U.S.A. 

tel: 631-273-9600 

toll free: 1-800-232-2670 

fax: 631-273-9659

 
 

www.adctoday.com 

[email protected]

              ADC (UK) Ltd. 
              Unit 6, PO14 1TH 
              United Kingdom

IB p/n 93-adcuffplus-00 rev 0 (2/17/2021)

                                                             Printed in the U.S.A.

Implied Warranty: 

Any implied warranty shall be limited in dura-

tionto the terms of this warranty and in no case beyond the original 
selling price (except where prohibited by law). This warranty gives 
you specific legal rights and you may have other rights which vary 
from state to state. 

For Australian Consumers: 

 Our goods come with guarantees that 

cannot be excluded under the Australian Consumer Law.  You are 
entitled to a replacement or refund for a major failure and compen-
sation for any other reasonably foreseeable loss or damage.  You 
are also entitled to have the goods repaired or replaced if the goods 
fail to be of acceptable quality and the failure does not amount to a 
major failure.

Warranty:

 

ADC warranty service extends to the original retail purchaser only 
and commences from the date of delivery. ADC warrants its prod-
ucts against defects in materials and workmanship under normal use 
and service.  Adcuff inflation system components (cuff, tubing, bulb, 
valves, and connectors) are warranted for three years. 

What Is Covered:

 Replacement of parts, and labor. 

What Is Not Covered: 

Transportation charges to ADC

®

. Damages 

caused by abuse, misuse, accident, or negligence. Incidental, 
special, or consequential damages. Some states do not allow the 
exclusion or limitation of incidental, special, or consequential 
damages, so this limitation may not apply to you. To Obtain 
Warranty Service: Send item(s) postage paid to ADC, Attn: Repair 
Dept., 55 Commerce Dr., Hauppauge, NY 11788. 
Please include your name and address, phone no., proof of 
purchase, and a brief note explaining the problem. 

 

Отзывы: