12.0 CALIBRATION
•
Occasionally the scale should be verified whether it is
weighing correctly by measuring to a known mass.
•
Zero the scale. Place the mass on the centre of the platform
and note the reading. Calibrate the scale, if necessary.
PROCEDURE
•
While in the normal weighing mode, press and hold the
[Tare/Zero]
key for 4 seconds.
•
The display will show “
CAL
” along with the last selected unit.
The unit can be changed by using the
[Unit]
key to calibrate
in kg or lb.
•
Press the
[Print/Hold]
key. The display will show “
L xx
”
where
xx
is the Calibration weight which is user-selectable.
•
Use
the
[Tare/Zero]
key to change the flashing digit and the
[Print/Hold]
key to move to the next digit.
•
Press the
[Unit]
key to confirm the calibration weight. The
digit stops flashing.
Note:
If the selected mass is less than 10% of the capacity of
the scale, an error message “
CALEr
” will be displayed and the
scale will return to zero. Repeat the process correctly.
© Adam Equipment Company 2008
15