©Adam Equipment Company 2008
16
8.1
FILTER
•
The value as set by the factory will be displayed (for ex. “
F L t 0
”).
•
Press the
[x10 /Inc]
key to scroll through the other options (
0
to
9
see
Note
below).
•
Select the desired setting by pressing
[Kg/lb /Set]
and move to the
next parameter for setting the communication parameters.
Note:
FLt 0
= Standard filter.
1
= Increased filter.
2 to 9
= High filter with hold for animal weighing.
Select the desired settings by pressing
[Kg/lb /Set]
and move to the next
parameter.
8.2
COMMUNICATION PARAMETERS
These set of parameters allow you to set the values for functions that
control communicating with a PC or a printer via the RS-232 interface.
8.2.1 COMMUNICATION ADDRESS
•
The value as set previously will be displayed (for ex. “
A d S
00
”).
•
If you wish to change it, press
[Acc /
].
The first digit will start
flashing.
•
Use the
[Acc /
]
key to advance to the next digit and the
[x10 /Inc]
key to increment the value of the digit.
Note:
AdS 00
= Continuous output mode.
01-98
= External command mode from PC
99
= Manual print mode with
[Acc /
]
Once the value is entered, press
[Kg/lb /Set]
to move to the next
parameter for setting the baud rate.
Store the desired settings by pressing
[Kg/lb /Set]
and move to the next
parameter