14.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an
error message. If an error message is shown, repeat the step that caused
the message. If the error is still shown contact your supplier for support.
ERROR
DESCRIPTION
POSSIBLE CAUSES
Err 4
Initial Zero is greater than
allowed when power is turned
on or the Zero key is pressed.
Weight is on the pan when
turning the scale on.
Excessive weight on the pan
when zeroing scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct when
turning the scale on.
Platform is not installed.
Load cell is damaged.
Electronics is damaged.