© Adam Equipment Company 2008
12
voltage will be offset by the amount of zero load voltage measured.
The load cell can be tested while it is connected to the A/D converter circuit board, using
the scale power supply for excitation voltage.
7.0 FACTORY PARAMETERS
The factory parameters are used for setting the capacity and basic settings.
Switch off the scale and then switch on again. Press
[Tare]
during the initial self-test.
The display will ask for a code number,
“
P _ _ _ _
“.
Enter the number 1000 and press
[Tare].
The display will show the first parameter
“
F1 CAL
”.
To scroll through other parameters
press the
[Limit]
key.
The display will show the parameter number then the word describing the function.
Press the
[Tare]
key to enter a parameter.
Press the
[Zero]
key to exit a parameter. Pressing
[Zero]
again will return the scale to
weighing.
When a parameter is entered by pressing the
[Tare]
key, the display will guide you
through the parameter selected and the options available.
The parameters available are:
F1
CAL
To enter the Calibration
F2
CAP
Select Capacity
F3
int
Initial Zero Range
F4
rE Z
Re-Zero Range
F5
SCS
Successive Tare Enable
F6
Cnt
Display the A/D counts
F7
Ldv
Set low voltage detect mode
F8
Uin
Calibration/parameter counter enabled
Содержание WBW 16
Страница 2: ... Adam Equipment Company 2008 ...
Страница 4: ... Adam Equipment Company 2008 2 ...
Страница 6: ... Adam Equipment Company 2008 4 ...
Страница 19: ... Adam Equipment Company 2008 17 8 0 WIRING DIAGRAM ...
Страница 20: ... Adam Equipment Company 2008 18 9 0 MECHANICAL ASSEMBLY ...
Страница 21: ... Adam Equipment Company 2008 19 ...