Load cell connecting definition (5pin aviation waterproof connector), see the area in the
circle:
1pin:
(
E+
)
red cable
2pin:
(
E-
)
black cable
3pin:
(
SH
)
shield cable
4pin:
(
S+
)
green cable
5pin:
(
S-
)
white cable
5.0 CALIBRATION
The PTS plus is factory calibrated, if further information is required for re-calibrating the
scale the procedure is as follows:
Turn on the power and wait for the scale to configure.
Before configuration is completed, hold down the
[Tare]
↵
key to bring up the
user password information screen.
To enter the password, use the directional buttons
⭡
or
⭣
to scroll through
numbers 1-9. Use the
⭢
directional key to move to the next digit.
Entering the correct password
[1000]
and pressing
[Tare]
for confirmation
will bring you to the user setup menu;
Within the menu, use the directional buttons
⭡
or
⭣
to scroll through the
settings until
‘U-cal’
appears on the display. Press the
[Tare]
↵
key to select.