The balance will now request the weight of the sample in air by displaying “
AIR WT
AIR WT
AIR WT
AIR WT
”. Place
the item on the pan, or in receptacle if the density kit is used, and press [Setup]. The
weight in air will briefly be shown in the last weighing unit selected.
After completion of the air weighing, the balance will request the weight in liquid by
displaying “
LIq wt
LIq wt
LIq wt
LIq wt
”. Submerge the item in the liquid and press [Setup] to start the liquid
weighing. The weight in liquid will briefly be shown in the last weighing unit selected,
followed by the computed density of the sample displayed as “XX.XXX
g/cc
”.
Remove the item from the pan and press [Mode] to continue with a new sample or press
[
0/T
] to return to normal weighing.
10.4.4.2
Density of a Liquid
When finding the density of a liquid, it is necessary to weigh a sample of known volume in
air and then in the liquid. The volume of the sample must be entered by the user. The last
known volume is stored for use at any time.
If using the density determination kit, the volume of the plumb is marked on its support,
e.g. 10.123 cc.
Steps:
Press [Mode] and then [Up] and [Down] keys until “
Density Liquid
” symbol is displayed and
then press [Setup] to enter this chosen density mode.
When “
Density Liquid
” is selected, the type of liquid used for the test must be selected:
The volume will be asked for by displaying “
VOLvME
VOLvME
VOLvME
VOLvME
” followed by a value which is the bulb
volume in cubic centilitres (cc). Enter or change the volume if required, using the keypad
numeric entry method (see section Error! Reference source not found.) and then press
[Setup] to continue.
The balance will now request the weight in air by displaying “
AIR WT
AIR WT
AIR WT
AIR WT
”. Place the glass
plumb supplied with the density determination kit in air on the weighing pan and press
[Setup] to start the air weighing. The value will briefly be shown in the last weighing unit
selected. The balance will now request the weight in liquid by displaying “
LIQ Wt
LIQ Wt
LIQ Wt
LIQ Wt
”.
Submerge the glass plumb in the liquid and press [Setup] to start the liquid weighing. The
weight will briefly be shown in the last selected unit, followed by the computed density of
the sample displayed as “XX.XXX
g/cc
”
Remove the item from the pan.
Press [Mode] to continue with a new sample or press [
0/T
] to return to normal
weighing.
If a printer or other serial device is connected then pressing [Print] will print the density
value in g/cc.
Содержание Nimbus NBL Serie
Страница 46: ...EN Adam Equipment Company 2015 44 3016612481Rev2 00 Jul15 18 BALANCE MENU STRUCTURE Operator Level Access ...
Страница 48: ...EN Adam Equipment Company 2015 46 3016612481Rev2 00 Jul15 Supervisor Level Access continued ...
Страница 150: ...DE Adam Equipment Company 2015 148 3016612481Rev2 00 Jul15 Deutsch ...
Страница 152: ...DE Adam Equipment Company 2015 150 3016612481Rev2 00 Jul15 Zugriff auf Supervisor Ebene Deutsch ...
Страница 153: ...DE Adam Equipment Company 2015 151 3016612481Rev2 00 Jul15 Zugriff auf Supervisor Ebene Fortsetz Englisch ...
Страница 154: ...DE Adam Equipment Company 2015 152 3016612481Rev2 00 Jul15 Zugriff auf Supervisor Ebene Fortsetz Deutsch ...
Страница 207: ...ES Adam Equipment Company 2015 205 3016612481Rev2 00 Jul15 Nivel de Acceso del Supervisor continuación ...
Страница 257: ...IT Adam Equipment Company 2015 255 3016612481Rev2 00 Jul15 18 STRUTTURA DEL MENU DELLA BILANCIA Operator Level Access ...
Страница 259: ...IT Adam Equipment Company 2015 257 3016612481Rev2 00 Jul15 Supervisor Level Access continued ...