5)
The
display
will
show
the
first
weight
that
you
can
use
to
calibrate
the
unit.
You
can
change
this
value
by
pressing
the
[Mode]
key.
The
weights
that
can
be
used
are
as
follows:
Model
#
HCB123
HCB153
HCB302
HCB602
HCB602H
HCB1002
HCB1502
HCB3001
Weight
1
60g
50g
100g
200g
200g
500g
500g
1000g
Weight
2
120g
100g
200g
400g
400g
1000g
1000g
2000g
Weight
3
‐
150g
300g
600g
600g
‐
1500g
3000g
6)
Once
the
calibration
weight
has
been
selected
press
the
[Tare]
key
8)
The
display
shows
LoAd.
Place
your
weight
on
the
pan.
Press
[Tare]
key
once
the
stable
sign
is
shown.
9)
The
display
will
show
PASS.
Remove
the
weight
from
the
pan.
NOTE:
If
the
calibration
fails
retry.
The
balance
will
show
FAlL H
(when
the
weight
is
higher)
or
FAlL L
(when
the
weight
is
lower).
Repeat
the
process
using
the
correct
calibration
weight.
21
|
P a g e
©
Adam
Equipment
Company
2008