4.0  CALIBRATION 

The  scale  can  be  calibrated  using  kilogram  weights  or  using  pounds  weights,(for  OIML  models,  pound  is 
disabled) depending on the weighing unit selected at the time of calibration. 
 

Steps: 

 
To  start  the  calibration,  turn  the  scale  off  and  switch  on  again  and  then  press

  [Tare]

  during  the  self-test.  

Enter code number 0000 and press

 [Tare]

. This will take you directly to the calibration section or you can get 

into the calibration section through the Indicator Settings (“FUnC 3”- see menu structure). 
 
The display will show 

"UnLoAd"  

 
Remove all weight from the platform and then press the

 [Tare] 

key when the display is stable. After the Zero 

point is set, the display will show 

“Ld  xx”

. Place the suggested calibration mass on the platform. It is best to 

use a weight close to the full capacity of the indicator.  
 
 If the mass is different from the displayed value, enter the value of the mass in whole numbers. The kg or 
the lb symbol will be on to show the active unit. Press the 

[Tare] 

key when the stable indicator is on.   

 
 When complete, it will display

 “PASS”

 and then either display 

“S8  CAL” 

(if entered the calibration section 

through the Scale Settings) or return to normal weighing (if entered directly). Remove the calibration mass. 
 
If an error message 

“FAIL H” 

or

 

“FAIL L” 

i

s shown, re-check the calibration and repeat. If the error cannot be 

corrected contact your supplier. 

 

4.1 

 

CALIBRATION

  

COUNTER

 

FOR

 

APPROVED

 

SCALES 

 

The approved (GK-M, GBK-M, GK-M Models) scales have the ability to control access to the calibration or 
metrology parameters using a passcode to limit access. The requirements for doing this stipulate the code 
should be apparent and recorded in a suitable location on the scale.   
 
At  power  on,  the  display  will  show  the  current  software  revision  number    followed  by  the  message  of  the 
Calibration Count   

“[AL[nt” 

then a number i.e.  “123”. The number is from the counter memory. Then the 

Parameter Counter message of

 “PAr[nt”

  and probably a different number.   The counters cannot be reset to 

0,  they  will  increment  until  the  display  can  no  longer  hold  the  values.    (1  to  999999).  The  scale  will  then 
continue to do the display test and go to normal weighing. 
 
If during the time the counting displays are shown, the user presses the

 [Tare] 

key, the user will be given a 

message to enter the passcode necessary to calibrate the scale, “P - - - - “   Enter the code  “P0000”  to 
Enter calibration or “P1000” to enter the parameters, followed by pressing the

 [Tare]

 key.   

The Calibration access will allow user calibration and the parameter code will allow access to the following 
parameters. 

 

Содержание GBK 120

Страница 1: ...ecycled materials and environmentally friendly packaging where possible As part of this initiative we have developed a short manual that uses less paper and ink to describe the main functions of your new Adam indicator scale A complete version is available at www adamequipment com Thank you for your support of Adam Equipment and we hope that you enjoy your new scale ...

Страница 2: ... Adam Equipment Company 2013 ...

Страница 3: ... 2 2 TARING 3 2 3 PARTS COUNTING 4 2 4 CHECK WEIGHING 5 2 5 PERCENT WEIGHING 6 2 6 ANIMAL Dynamic WEIGHING 7 2 7 ACCUMULATED TOTAL 8 3 0 RS 232 SPECIFICATION 9 4 0 CALIBRATION 10 4 1 CALIBRATION COUNTER FOR APPROVED SCALES 10 5 0 SPECIFICATIONS 11 6 0 ERROR MESSAGES 16 7 0 MENU STRUCTURE 17 8 0 WARRANTY INFORMATION 18 ...

Страница 4: ...P a g e 2 Adam Equipment Company 2013 ...

Страница 5: ...ght disabled the life is approximately 70 hours before needing to be recharged The battery should be charged for 12 hours for full capacity 2 0 OPERATION 2 1 ZEROING You can press the Zero key at any time to set the zero point from which all other weighing and counting is measured The scales have an automatic re zeroing function to account for minor drifting or accumulation of material on a connec...

Страница 6: ...tering parts counting and press Tare Press Cnt to enter the Parts Counting mode The display will show the last sample size used For example 10 Pcs To change the sample size you can press CE to clear the last values and then enter the value 20 using the numeric keypad Place the right number of parts on the platform Then press Cnt to determine an average piece weight After the sample has been weighe...

Страница 7: ... value will return the scale to weighing with the Check weighing function enabled During parts counting and percent weighing the limits are set in the same way as above The limits are displayed in pcs or To disable the check weighing function enter zero into both the limits as described above When the current limits are shown press CE to clear the settings then press Tare to store the zero values ...

Страница 8: ...ey F4 PCt will be displayed Press Tare again to enter percent weighing The scale will set the sample mass on the platform as 100 reference weight NOTE If there is no reference weight on the pan and percent weighing function is entered pressing Tare again will return the indicator to normal weighing Remove the sample weight Then any other weight placed on the platform will be displayed as a percent...

Страница 9: ...ll show Filt x where x is a value from 1 to 5 To increment the value shown press the Func key then press the Tare key to accept it The display will flash Ani 2 times then show the current weight 0 00 The scale is now ready to weigh Place containers or blankets onto the platform and press the Tare key to zero the display Place the animal to be weighed on the platform When a stable reading is found ...

Страница 10: ... 2 for 2 seconds and then show the new total Continue until all weights have been added This can continue for up to 99 entries until the capacity of display is exceeded To view the total in memory press the Print key when the indicator is at zero The display will show the total number of accumulation ACC xx and the total weight before returning to zero To print the total press Print to recall and ...

Страница 11: ...rface as standard The indicator when connected to a printer or computer outputs the weight with the selected weighing unit through the RS 232 interface Default Specifications Connector RS 232 output of weighing data ASCII code 9600 Baud user selectable 8 data bits No Parity 9 pin d sub miniature socket Pin 3 Output Pin 2 Input Pin 5 Signal Ground ...

Страница 12: ...rn to normal weighing if entered directly Remove the calibration mass If an error message FAIL H or FAIL L is shown re check the calibration and repeat If the error cannot be corrected contact your supplier 4 1 CALIBRATION COUNTER FOR APPROVED SCALES The approved GK M GBK M GK M Models scales have the ability to control access to the calibration or metrology parameters using a passcode to limit ac...

Страница 13: ...rating Temperature 10 C 40 C 14 F 104 F Power supply 230 VAC 50 60 Hz 12V 800ma adaptor for USA versions Battery Internal rechargeable battery Calibration Automatic External Display 6 digits LCD digital displays with capacity tracker and symbols for units Indicator Housing ABS Plastic Overall Dimensions wxdxh 260 x 170 x 115mm 10 2 x 6 7 x 4 5 Net Weight 1 8 kg 4 lb Applications Weighing and check...

Страница 14: ...for USA Calibration External Calibration Mass User Selectable Display Backlit Green display 20mm with capacity tracker Balance Housing Cast aluminum base Pantone cool grey painted base stainless steel grade 304 Top pan ABS Cool grey indicator housing Pan Size 400mm x 500mm x 65mm 15 7 x 19 7 x 2 6 400mm x 500mm x 65mm 15 7 x 19 7 x 2 6 400mm x 500mm x 65mm 15 7 x 19 7 x 2 6 600mm x 800mm x 65mm 23...

Страница 15: ...a only XXXa also to have pounds Ounces Pound Ounces Stabilization Time 2 3 Secs Operating Temperature 10 C to 40 C 14 F to 104 F Power Supply 230VAC 50 60Hz in Europe Asia and South Africa 12vDC 800mA UL CSA adapter for USA Calibration External Calibration Mass User Selectable Display Backlit Green display 20mm with capacity tracker Balance Housing Cast aluminum base Pantone cool grey painted base...

Страница 16: ...0VAC 50 60Hz Calibration External Calibration Mass User Selectable Display Backlit Green display 25mm with capacity tracker Balance Housing Cast Aluminum base Pantone cool grey painted base stainless steel grade 304 Top pan ABS Cool grey indicator housing Pan Size 400mm x 500mm 15 7 x 19 7 Overal Dimensions w x d x h 400 mm x 620 mm x 860 mm 15 7 x 24 4 x 33 9 Net Weight 12 5kg 27 6 Lb Features We...

Страница 17: ...F Power Supply 230VAC 50 60Hz Calibration External Calibration Mass User Selectable Display Backlit Green display 25mm with capacity tracker Balance Housing Cast Aluminum base Pantone cool grey painted base stainless steel grade 304 Top pan ABS Cool grey indicator housing Pan Size 300mm x 400mm x 50mm 11 8 x 15 5 x 2 Overal Dimensions w x d x h 300mmx 520mm x 660mm 11 8 x 20 5 x 26 Net Weight 7 6k...

Страница 18: ...t on the platform when turning the indicator on Excessive weight on the platform when zeroing the indicator Platform is not installed Improper calibration of the indicator Damaged load cell Damaged Electronics Err 6 A D count is not correct when turning the indicator on Load cell is damaged Electronics is damaged Err 7 Percent input error Percent function is entered with no reference mass on the p...

Страница 19: ... Prt on Prt oFF C2 bd Baud Rate 600 To 19200 C3 Prm Printing Mode mA StA Manual Stable mA AnY Manual Any Au StA Auto Stable Au End Auto End Ct StA continuous Stable Ct AnY Continuous Any C4 Aon Enable Accumulation on oFF C5 Ln Language for printing English French German Spanish C6 Uid User ID Enter using numeric keys C7 Sid Scale ID Enter using numeric keys C8 LAb LAb On Lab Off FUNC 3 Scale Param...

Страница 20: ...ginal packaging and with correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performance is due to misuse accidental damage exposure to radioactive or corrosive materials negligence faulty installation unauthorised modifications or attempted repair or failure to observe the requiremen...

Страница 21: ...19 P a g e Adam Equipment Company 2013 ...

Страница 22: ...commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense CALI...

Страница 23: ...Phone 49 0 4340 40300 0 Fax 49 0 4340 40300 20 e mail vertrieb aeadam de Adam Equipment S A Pty Ltd 7 Megawatt Road Spartan EXT 22 Kempton Park Johannesburg Republic of South Africa Phone 27 0 11 974 9745 Fax 27 0 11 392 2587 e mail sales adamequipment co za Adam Equipment S E ASIA PTY Ltd 2 71 Tacoma Circuit CANNING VALE 6155 Perth Western Australia Phone 61 0 8 6461 6236 Fax 61 0 8 9456 4462 e m...

Отзывы: