P a g e
10
© Adam Equipment Company 2013
4.0 CALIBRATION
The scale can be calibrated using kilogram weights or using pounds weights,(for OIML models, pound is
disabled) depending on the weighing unit selected at the time of calibration.
Steps:
To start the calibration, turn the scale off and switch on again and then press
[Tare]
during the self-test.
Enter code number 0000 and press
[Tare]
. This will take you directly to the calibration section or you can get
into the calibration section through the Indicator Settings (“FUnC 3”- see menu structure).
The display will show
"UnLoAd"
Remove all weight from the platform and then press the
[Tare]
key when the display is stable. After the Zero
point is set, the display will show
“Ld xx”
. Place the suggested calibration mass on the platform. It is best to
use a weight close to the full capacity of the indicator.
If the mass is different from the displayed value, enter the value of the mass in whole numbers. The kg or
the lb symbol will be on to show the active unit. Press the
[Tare]
key when the stable indicator is on.
When complete, it will display
“PASS”
and then either display
“S8 CAL”
(if entered the calibration section
through the Scale Settings) or return to normal weighing (if entered directly). Remove the calibration mass.
If an error message
“FAIL H”
or
“FAIL L”
i
s shown, re-check the calibration and repeat. If the error cannot be
corrected contact your supplier.
4.1
CALIBRATION
COUNTER
FOR
APPROVED
SCALES
The approved (GK-M, GBK-M, GK-M Models) scales have the ability to control access to the calibration or
metrology parameters using a passcode to limit access. The requirements for doing this stipulate the code
should be apparent and recorded in a suitable location on the scale.
At power on, the display will show the current software revision number followed by the message of the
Calibration Count
“[AL[nt”
then a number i.e. “123”. The number is from the counter memory. Then the
Parameter Counter message of
“PAr[nt”
and probably a different number. The counters cannot be reset to
0, they will increment until the display can no longer hold the values. (1 to 999999). The scale will then
continue to do the display test and go to normal weighing.
If during the time the counting displays are shown, the user presses the
[Tare]
key, the user will be given a
message to enter the passcode necessary to calibrate the scale, “P - - - - “ Enter the code “P0000” to
Enter calibration or “P1000” to enter the parameters, followed by pressing the
[Tare]
key.
The Calibration access will allow user calibration and the parameter code will allow access to the following
parameters.