41 |
P a g e
© Adam Equipment Company 2009
6.2
OTHER
SETTINGS
6.2.1 Restore Factory Calibration
•
To restore this calibration value it is necessary to do a simple
procedure. Press and hold the
[Zero]
key down for 3 seconds to put
the scale into a weighing display.
•
Press the
[Total]
key, the scale will beeps. The Factory Calibration
will be restored to the balance.
6.2.2
Shortcut to Calibration without using Functions
•
To calibrate the scale without entering the function menu press
and hold the
[MC/
]
key while power is turned on. The scale will
immediately go to the calibration function.
•
The display will show the following:
“
CAL
” “
0.000
” “
12345
”
•
Proceed with calibration as normal. See section 5.4.4. When
completed the scale will return to normal operation. You cannot
proceed to any other functions from this entry point.