8.9
F9
–LOW VOLTAGE DETECTION
This parameter allows detection of low voltage when the battery wears down.
To enter this parameter, press the
[Tare]
key when “
F9 Ld M
” is shown.
The display will show if the LVD Mode is set to
on
or
oFF
.
Press the
[Func]
key to scroll through the options.
Press
[Tare]
to accept the displayed value and return to the parameter or press the
[Zero]
key to return to the parameter without changing the value.
Then press
[Func]
to advance to the first parameter.
Pressing
[Zero]
will display “
JP on
” indicating the jumper is still in place. Switch off the
scale and remove the jumper.
NOTE : TO SECURE THE SCALE, IT IS NECESSARY TO REMOVE THE JUMPER AND
RE-SEAL THE SCALE.
NOTE:
The scale has a special feature for displaying x10 the value of the weight which
is a useful tool during testing. Press the
[1]
key for more than 3 seconds. It will show
one extra decimal place making it a high resolution display. Pressing the
[1]
key again
for more than 3 seconds will bring the display back to the normal resolution.