© Adam Equipment Company 2019
8.2
CALIBRATION
The calibration display should look as follows:
To calibrate the base, make sure the load cell cable is plugged into the back of the
indicator, where it is labelled up load cell, the scale base is level, and there is no
weight on the top pan.
Enter the weight you wish to calibrate the scale using the Green keys; you can use
the
abc
button to add a decimal point to the calibration value. Pressing the
[CE/backlight]
button clears any entered value. Press the
PLU
key to show the
zero A/D counts; when the stable indicator is showing press the
PLU
key to accept
the zero point.
The screen should look similar as follows:
Load the weight you wish to calibrate with on the top pan, and press the
PLU
key.
You should see Pass on the display, and then the indicator will return to normal
weighing.
NOTE:
BCT scales will have the lb or kg indicator on, to show the unit of the weight
requested. If the scale was in pounds before starting the calibration, the weights requested
will be in pound values or if the scale was weighing in kilograms then metric weights will be