8.0 CALIBRATION
8.1
SIMPLE
CALIBRATION
The following steps will allow the user to conduct the simple calibration procedure.
Power off the indicator.
Press and hold [Q’TY PST] key, then power on.
The weight window will show the version number, total quantity window will show -
CAL-, about 2 second later, alarm will beep once.
Input the value of the weights, put on the right weights, after the stable sign
appeared, press [Q’TY PST], and then calibration succeed.
For more specific details about how to calibrate your indicator, please contact the
manufacturer directly.
NOTE:
AE504 indicators will have the lb or kg indicator on, to show the unit of the weight
requested. If the scale was in pounds before starting the calibration, the weights requested
will be in pound values or if the scale was weighing in kilograms then metric weights will be
requested.
9.0
LABEL
PRINTING
FUNCTION
In normal weighing mode, press
[Print/Save]
to print through built in label printer.
Printing can also be done through RS 232 connectivity.
Labels can also be printed via the built-in printer when in the accumulation recall
function and in parts counting.
9.1
LABEL
EDITING
COMPUTER
SOFTWARE
The AE 504 is supplied with RS-232 interface which allows for connection between
computer and indicator to edit labels for the inbuilt label printer and set printing
parameters.