11.0 CALIBRATION
The AE 403 scales calibrate using either metric or pound weights, depending on the
weighing unit in use before calibration. The display will show either "kg" or "lb" to identify
the weights expected.
The scale can be calibrated using the following procedure:
Turn on the power and wait for the scale to configure.
Before configuration is completed, hold down the
[Tare]
↵
key to bring up the
user password information screen.
To enter the password, use the directional buttons
⭡
or
⭣
to scroll through
numbers 1-9. Use the
⭢
directional key to move to the next digit.
Entering the correct password
[1000]
will bring you to the user setup menu.
Within the menu, use the directional buttons
⭡
or
⭣
to scroll through the
settings until
‘U-cal’
appears on the display. Press the
[Tare]
↵
key to select.
After pressing
[Tare]
↵
,
‘noload’
should appear on the display. Use the
directional button
⭡
to switch to
‘load1’
and set the desired weight limit of the
test object to be placed on the scale using the
⭡
or
⭣
directional buttons and
the
⭢
directional key to move to the next digit. Press the
[Tare]
↵
key to
confirm.
After pressing
[Tare]
↵
,
‘Load’
will appear on the display. Place the
calibration test weight that matches the weight previously entered onto the
scale weighing pan and press the
[Tare]
↵
key to confirm.
Pressing Tare should bring up the
‘Load 2’
option on the display. Follow the
same procedure as before by setting a new weight and adding the
appropriate calibration weight and press the
[Tare]
↵
key to confirm.
Pressing
[Tare]
↵
will reconfigure the scale and bring the user back to the
regular weighing mode.