Main Menu
All functions of the SpectraPro-300i are accessed while the controller displays the Main Menu.
GOTO Menu:
Press the white key below the GOTO command. The following menu will appear (with a flashing arrow next
to the
Enter Wavelength
line):
Key in the desired wavelength (in nm) and then push the <ENTER> key (ex: 500 <ENTER>). The
SpectraPro-300i will then race at the maximum motor speed to the selected wavelength and display this
interim menu.
GOTO Interim Menu
11
Acton Research Corp.
SP-300
Wavelength=
0.0nm
Scan to
0.0nm
Rate=
100.0 nm/min
Grating #1
1200 g/mm
Jog Incr
1.00nm
GOTO
SCAN
SPEED
GRAT
TURR
QUIT
Note: When entering a number for GOTO, SCAN or SPEED, the controller
will accept an integer or a number to the nearest tenth of a
nanometer (nm). If you enter a number with hundredths or lower, it
rounds up to the nearest tenth nanometer.
{ HOME }
PRG
Go To:
Enter Wavelength
Acton Research Corp.
SP-300
Wavelength=
234.5nm
Scan to
500.0nm
Rate=
100.0 nm/min
Содержание SpectraPro-300i
Страница 24: ...Appendix V C Standard SpectraPro Slit Assembly Drawing Slit Position 23...
Страница 32: ......