Operating Instructions

Acton Research Corporation

SpectraPro-300i

0.300 Meter Focal Length Triple Grating

Imaging Monochromator / Spectrograph

 

SP-300i Manual Rev. 997.1

 

530 Main Street, Acton, MA  01720 

      Phone: 

(978)263-3584, 

Fax: 

(978)263-5086

 

Web Site: 

www.acton-research.com

Содержание SpectraPro-300i

Страница 1: ...earch Corporation SpectraPro 300i 0 300 Meter Focal Length Triple Grating Imaging Monochromator Spectrograph SP 300i Manual Rev 997 1 530 Main Street Acton MA 01720 Phone 978 263 3584 Fax 978 263 5086...

Страница 2: ...716 Bilateral Slits and 715 Fixed Slit Assemblies 8 9 IV SpectraPro 300i Operation A Initialization 10 B Operating the SpectraPro 300i Using the Model 300 749 Remote Scan Controller 10 14 C Operating...

Страница 3: ...atism corrected optical system with multichannel imaging capabilities Integrated microstepped stepping motor with controller controlled by a powerful 32 bit microprocessor Built in RS232 and IEEE488 c...

Страница 4: ...e Less than 20 seconds Grating Drive System 32 bit microprocessor controlled direct digital scanning DDS system Scan Linearity The SpectraPro 300i scans linear with respect to wavelength Size 13 25 lo...

Страница 5: ...onnections necessary for operation from a computer through the COM ports or with a GPIB controller card Figure 2 below shows the cable connections necessary for operation of the SpectraPro 300i with t...

Страница 6: ...optional If none of these cables are compatible with your system consult ARC for a custom cable If you have facilities for constructing a custom cable use the RS232 pin arrangement shown in Table 1 o...

Страница 7: ...flange for CCDs and diode arrays accommodates detectors with two different bolt circles There are three 3 equally spaced 10 32 tapped holes on a 3 60 bolt circle and three 3 equally spaced holes on a...

Страница 8: ...the spacer if not required 4 Position the array detector mounting flange against the array detector and match the hole patterns When correctly positioned the baffle aperture orientation should match t...

Страница 9: ...top of the slit housing The micrometer knob is graduated in 0 010 millimeter 10 m increments Each clockwise revolution of the micrometer knob increases the slit width 0 25 millimeters 250 m For maximu...

Страница 10: ...ve the slit plate without contacting the slit blades Store the removed slit plate in a clean dry area 3 Place the new slit plate in the slit body and secure it in position using the slit plate screws...

Страница 11: ...roller is connected to the SpectraPro 300i through the standard RS232 port with the cable provided It is designed to send commands and receive information about the specific SpectraPro 300i it is atta...

Страница 12: ...tor speed to the selected wavelength and display this interim menu GOTO Interim Menu 11 Acton Research Corp SP 300 Wavelength 0 0nm Scan to 0 0nm Rate 100 0 nm min Grating 1 1200 g mm Jog Incr 1 00nm...

Страница 13: ...n the desired wavelength in nm followed by ENTER The SpectraPro 300i will scan to the selected wavelength at the preset scanning rate To change the scanning rate see the SPEED Scanning Rate instructio...

Страница 14: ...your SpectraPro 300i From the Main Menu press the white key below the GRAT command This will display a menu that shows the available gratings for the turret currently installed with a dot indicating...

Страница 15: ...ns The controller for the SpectraPro 300i allows the user to change wavelengths without accessing GOTO or SCAN and without keying in a destination wavelength This function is useful for rapidly scanni...

Страница 16: ...o complete the processing of that command string before sending another command This is accomplished by checking the status byte The SpectraPro controls bits in the IEEE488 status byte which can be re...

Страница 17: ...to nine 9 gratings are allowed on three 3 turrets This command takes a grating number from 1 9 IMPORTANT NOTE This command assumes that the correct turret is specified by the TURRET command For exampl...

Страница 18: ...the diverter control commands This command is for SpectraPro monochromators which can accept two diverter mirrors The SP 300i monochromators will not accept this command FRONT Moves the designated di...

Страница 19: ...is to maintain compatibility with previous SpectraPro applications For new applications use the INIT SP300 GADJUST command below INIT SP300 GADJUSTSets grating adjustment value for the designated gra...

Страница 20: ...ce 3 With the array detector operating check the image of the light source if running in an imaging mode with a CCD Otherwise check the line intensity and shape 4 Using a 1 8 allen wrench loosen the t...

Страница 21: ...gnment and focus of array detectors Spectrograph to Monochromator This procedure requires that you have the spectrograph version Model SP 306 plus the Model SP 716 2 exit slit assembly 1 Loosen the tw...

Страница 22: ...ult INIT WAVELENGTH Sets an initial wavelength for the SpectraPro 300i after initialization E G 435 84 INIT WAVELENGTH Notice that two digits after the decimal point are required INIT SRATE Sets an in...

Страница 23: ...300i to PC AT or Compatible SpectraPro 300i to PC XT or Terminal SpectraPro 300i to Macintosh Computer Modem or Printer Port DB9P MALE DB9P MALE DB9P MALE 8 PIN MALE 1 2 3 4 5 6 7 8 1 6 2 7 3 8 4 9 5...

Страница 24: ...Appendix V C Standard SpectraPro Slit Assembly Drawing Slit Position 23...

Страница 25: ...180 optical paths SP 308 SpectraPro 300i 300mm Monochromator Spectrograph Same as SP 305 except with micrometer controlled entrance and exit slits 180 apart motorized port selection mirror and multich...

Страница 26: ...slits and a motorized exit slit selection mirror Provides 90 and 180 optical paths SP 558 SpectraPro 500i 500mm Monochromator SpectroGraph Same as SP 555 with micrometer controlled entrance and exit...

Страница 27: ...indow PMT and photometer power supply for operation Accepts ARC tubes P1 P2 or P3 below or customer specified 1 1 8 side window PMTs PD 439 Same as PD 438 above also includes integrated light tight sh...

Страница 28: ...ssembly Includes 320nm 590nm 665nm and 715nm filters mounted in FA 448 2 filter wheel Enables computer controlled indexing of order sorting filters for elimination of unwanted second order radiation R...

Страница 29: ...IR Fiber Optic Bundle 3 0 meter long for 400nm to 2200nm Requires mount FC 446 020 Slit mount with horizontal translation capability for 10mm diamter fiber optic bundle Connecting Cables CC 499 1 Cabl...

Страница 30: ...y fax with written confirmation Acton Research Corporation Please include complete name and address with these orders P O Box 2215 FAX 978 263 5086 Acton MA 01720 U S A Orders accepted subject to curr...

Страница 31: ...fect appears within the warranty period the purchaser shall promptly notify ARC No material will be accepted for repair or replacement without prior authorization from ARC Upon such authorization and...

Страница 32: ......