5.13.1. SUGGESTED INCLINATION ANGLES
MAXILLA
Molar
35°
Premolar and canine
45°
Incisor
55°
Bite-wing exam
10°
-----------------------------------------------------------
Bite-wing exam
0°
Incisor
-20°
Premolar and canine
-10°
Molar
-5°
MANDIBLE
5.14. MAKE THE EXPOSURE
PLEASE NOTE
To obtain the best radiological diagnostic results, pay extreme attention in all the steps of the process: positioning the patient, the x-ray
generator and collimator, exposing the x-ray detector (film/digital sensor/plates); and processing the film or the digital image acquisition
and processing chain.
CAUTION
It is mandatory for the operator to check the parameters before making any exposure with either the local or remote exposure switch.
CAUTION
Always maintain audio and visual contact with the patient and unit during the x-ray exposure: the operator is required to monitor the
patient and the X-Mind unity parameters throughout the entire duration of the x-ray examination.