6. Troubleshooting
Fault Possible
cause Solution
Unplugged or defective mains cord.
Check and / or replace the mains cord.
Fuses blown.
Replace the fuses.
Defective wall power outlet.
Contact your electrician.
Device switched off (0).
Set the switch to the on position (I).
Does not operate.
(general)
Foot control disconnected and / or defective.
Check and / or replace the foot control.
No air or water coming from the pipes connected to the
mains supply.
Contact your approved installer.
No powder, air or water.
Handpiece and / or nozzle blocked.
Disconnect the handpiece and check that air and powder are coming out of the
cord:
- if not, return the device to Satelec After Sales Service;
- if they are, unscrew the handpiece nozzle, unblock the nozzle with a file and
place the handpiece in an ultrasound tank for 10 minutes.
Powder: drum empty or too full.
Fill the drum or remove the excess.
Water: water inlet not connected or incorrect water setting.
Check the water inlet pipe or adjust the water flow on the handpiece cord.
Powder damp.
Empty and clean the drum and refill with new powder.
No powder or water but air
present.
Handpiece and / or nozzle blocked.
Disconnect the handpiece and check if water comes out of the cord:
If it doesn’t, return the device to Satelec After Sales Service.
If it does, unscrew the handpiece nozzle, unblock the nozzle with a file and place
the handpiece in an ultrasound tank for 10 minutes.