Possible causes
Solutions
Angle of approach incorrect or inadequate pressure
on the clinical site
Refer to the tip user manual indicated
Moisture on the I-Surge LED micromotor or handpiece
cord connector
Dry the electrical contacts
10.2.3 No spray or very little amount of spray
Symptoms: when the device is in use, irrigation is not working and no spray comes out of the contra-angle or
the tip
Possible causes
Solutions
Irrigation solution bag empty
Replace the irrigation bag with a new one
Irrigation deactivated
Start irrigation using the footswitch or touch-sensitive screen
Irrigation line blocked
Replace the irrigation line
Irrigation line pinched
Check the length of the irrigation line
Incorrect irrigation adjustment
Adjust the irrigation flow rate
10.2.4 Light not working or too dim
Symptoms: the handpiece is not lighting up the clinical site
Possible causes
Solutions
LED ring missing in the handpiece nozzle
Install the LED ring
Faulty LED ring
Replace the LED ring
Cracks or fine cracks on the LED ring
Replace the LED ring
LED ring contacts faulty
Replace the LED ring
LED ring poles reversed
Install the LED ring with the correct polarity
Optical guide damaged
Replace the optical guide
Handpiece cord connector faulty
Dry the cord and medical device connectors
Replace the handpiece and cord
Other
Return the medical device to the SATELEC
®
customer service team
10.2.5 Water leakage
Symptoms: water is leaking from one of the following places:
l
irrigation cassette
l
handpiece
Possible causes
Solutions
The handpiece irrigation inlet is damaged
Replace the handpiece and cord
Return the handpiece to the SATELEC
®
customer service team
Pipe ruptured in the irrigation line cassette
Replace the irrigation line
Irrigation cassette not working
Replace the irrigation line
10.2.6 Ultrasounds not working
Symptoms: the tip does not vibrate, vibration cannot be heard.
Possible causes
Solutions
Tip loose
Fasten the tip using the wrench
Connector with the medical device
faulty
Clean the cord contacts
Handpiece cord wire(s) cut
Return the handpiece and cord to the SATELEC
®
customer service team to have
them replaced
User Manual • ImplantCenter™ M+ • J27251 • V1 • (13) • 09/2013 • NO25EN010A - Page
29
/
42
Содержание ImplantCenter M+
Страница 1: ...User Manual ImplantCenter M...
Страница 2: ......
Страница 5: ...15 Glossary 41...
Страница 6: ......
Страница 8: ......
Страница 26: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 24 42...
Страница 28: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 26 42...
Страница 40: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 38 42...