Possible causes
Solutions
The lid O-ring is incorrectly
positioned
There are traces of powder under
the lid O-ring
The lid O-ring is defective
The tank is cracked
l
Unscrew the lid from the tank;
l
Pour the powder into an infectious clinical waste container;
l
Remove and inspect the lid O-ring;
l
Using the multi-purpose syringe air function, blow to clean the lid
threads;
l
Install the lid O-ring;
l
Fill the tank up to the indicated limit with the powder required for
the treatment to be performed;
l
Screw the lid back onto the tank.
8.4 Water in the powder tank
Symptoms: drops of water appear in the powder tank.
Possible causes
Solutions
The tank wasn't dry when the powder
for poured in
l
Unscrew the lid from the tank;
l
Pour the powder into an infectious clinical waste container;
l
Purge the air circuit at least three times for five seconds;
l
Dry the air circuit using the multi-purpose syringe air function;
l
Wipe the inside walls of the tank;
l
Fill the tank to the indicated limit with the powder required for
the treatment to be performed;
l
Screw the lid back onto the tank.
The quick coupling O-ring is defective
The malfunction is with the dental chair. Contact a technician.
There is water in your compressor
Contact a technician to check your compressor.
User Manual • AIR-N-GO® easy • J10121 • V5 • 11 • 10/2015 • ND27EN050E - Page
23
/
35
Содержание AIR-N-GO easy
Страница 1: ...User Manual AIR N GO easy...
Страница 6: ......
Страница 8: ...Page 6 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 16: ...Page 14 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 26: ...Page 24 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 28: ...Page 26 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 32: ...Page 30 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 34: ...Page 32 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...