background image

Silicon Explorer Quick Start

4

Connecting the 
Probe Leads

You can connect the probe leads by 
attaching one of two supplied cable 
assemblies.

One cable assembly terminates in 
individual 0.025 receptacles that can 
connect to 0.025 headers or micro-clips 
according to the labels on the wires — 
the other assembly interfaces directly to 
the target board when a 16-pin header 
is installed on the target according to 
the pin-out shown in Figure 1.

The Probe A and B pins on the probe 
connector may be selected as the source 
for logic analyzer channels 1 and 2 
respectively, by clicking on the # field in the waveform window 
and selecting PRA or PRB as the source. Selecting the probe pins 
as the source disables the CH1 and/or CH2 inputs from the 
target cable assembly.

Using the Probe Instrument

The following procedures describe how to use the Probe 
Instrument.

To open a Probe file:

1.

Launch The Silicon Explorer software. 

Select Silicon 

Explorer from the Designer menu under Programs in the Start 
menu.

2.

Select a Probe (.prb)

 

file to open. 

Select the Open 

command from the File menu. Choose (*.prb) as the Files of 
type and double click the desired probe file.

The Probe Instrument uses a .prb file during analysis, which 
is exported from the Designer software. This file contains 
device information, net names, and node location 
information. This information is displayed in a tree structure 
for verification by the user on the left side of the Analyze 
window after a Probe file is opened.

 2

 4

 6

 8

10

12

14

16

 1

 3

 5

 7

 9

11

13

15

Rsvd

Rsvd

Rsvd

Mode

Rsvd

GND

GND

GND

SDI/TDI

DCLK/TCLK

SDO/TDO

TMS

Rsvd

VCC

PRA

PRB

PCB Layout

Component Side

Figure 1. Pinouts

Содержание Silicon Explorer

Страница 1: ...Silicon Explorer Quick Start Windows Environment...

Страница 2: ...icular purpose Informa tion in this document is subject to change without notice Actel assumes no responsibility for any errors that may appear in this document This document contains confidential pro...

Страница 3: ...e Target 2 Powering the Silicon Explorer 2 Current Consumption 2 Using an External Power Supply 3 Connecting the Probe Pilot to the FPGA 3 Connecting Signals to the Target for Logic Analysis 3 Connect...

Страница 4: ......

Страница 5: ...Silicon Explorer A more detailed description can be found in the on line help system The system requirements for Silicon Explorer and Silicon Explorer Lite are Designer R3 1998 or later 486 66 microp...

Страница 6: ......

Страница 7: ...necting to the PC The Probe Pilot and Probe Pilot Lite connect to a standard serial port using either an 8250 or 16550 UART Communication is established via the 9 Pin D Sub connector Software Settings...

Страница 8: ...rer is designed to withstand the abuse normally found in a lab environment However long term exposure to out of range conditions can cause failure In particular long term connection to reverse or over...

Страница 9: ...a 22 pin 18 channels a clock VCC GND and clock GND and a 16 pin connector for FPGA debug using Actel s Action Probe circuitry The Probe Pilot Lite only has the 16 pin connector Connecting Signals to...

Страница 10: ...CH2 inputs from the target cable assembly Using the Probe Instrument The following procedures describe how to use the Probe Instrument To open a Probe file 1 Launch The Silicon Explorer software Selec...

Страница 11: ...the debug connector The net name is also displayed in the waveform window Using the Timing Instrument Note This section does not apply to Silicon Explorer Lite Select Silicon Explorer from the Design...

Страница 12: ...is recognized or the stop button black square is pressed see Figure 2 During this period the status window displays PRE for the pre trigger state although you may not see it if the trigger pattern is...

Страница 13: ...requirement is fulfilled Then the data are uploaded to the host READ State with progress indicator The entire upload process takes less than 15 seconds although you may zoom in and view data immediat...

Страница 14: ...t features for viewing saving and printing data More detailed information is available from the on line help system Troubleshooting FAQs Q1 I ve connected Probe Pilot and the Silicon Explorer window d...

Страница 15: ...Silicon Explorer window flashes an error message and returns to IDLE A If the target power is incapable of supplying the necessary current then the Silicon Explorer window resets and returns to IDLE M...

Страница 16: ...ng test 1 Connect the supplied ribbon cable from the Probe Pilot or Probe Pilot Lite to the test header and apply power to the terminals on the test header Red is positive Black is negative 2 Disconne...

Страница 17: ...408 739 1010 Lutyens Close Lychpit 1 24 14 Ebisu Shibuya ku Customer Service FAX 408 522 8044 Basingstoke Toyko 150 Japan Action Facts Technical Documents 800 262 1062 Hampshire RG24 8AG Tel 81 0 3 34...

Отзывы: