![Actegy REVITIVE 2836AA Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/actegy/revitive-2836aa/revitive-2836aa_user-manual_2842285016.webp)
30
31
Problem
Possible Cause
Solution
When using the
Arthritis Pads, my
thigh muscles are
aching before the end
of the treatment.
Your muscles may not yet be
strong enough to carry out a full
20 minute session.
Turn the device off and end the
session before the end of the 20
minutes. On your next session
start on a lower setting (where
you can feel the mild electrical
stimulation and it is comfortable)
and reduce the duration until
your muscles have acclimatized
to the stimulation.
When using the
Arthritis Pads, I can’t
feel the stimulation in
my thigh.
Therapy has started and has
reached the rest period.
You will need to wait for up to 50
seconds before the stimulation
will start again.
For more information including Frequently Asked Questions,
please visit www.revitive.com
Name of product
REVITIVE Arthritis-Knee
Model
2836AA
Frequency
1Hz – 50Hz
Output current
Max 15mA
Weight
1.5kg
Dimensions (with IsoRocker enabled and disabled)
enabled: 355(W)x355(H)x75(D) mm
disabled: 355(W)x355(H)x90(D) mm
Power consumption
5W
AC adaptor
CE Approved
Power source
Input (adaptor used)
Output
100-240V AC ( ), 50/60Hz, 0.18A
5V ( ) DC ,1.0A
Battery & Battery Life
Li-Ion / 500 recharge cycles
Contraindications
This describes situations where you should not use the
REVITIVE
Warnings and Cautions
Make sure you understand these before using the
REVITIVE
Power
Time Remaining
Intensity Level
Battery Indicator
IsoRocker Disabler
Center Positive Polarity
Type BF medical electrical equipment
Class II medical electrical equipment double insulated
Technical Specifications
Troubleshooting
2910_IFU02_12517049.indd 30-31
22/07/2016 09:49