7.4 FUNCTIONAL CHECK
Pay attention to ease of movement when releasing the plug connection. Exaggerated application of
a force can damage the plug connection. Avoid bending or levering on the plug connections.
The functional check must be carried out before each measu-
ring task.
1. Apply some couplant to the probe.
2. Calibration block and probe must be mutually connected, by insertion into the bay-
onet connector and careful turning.
3. Switch ON the device.
4. The value displayed must be within the range of the values specified on the cali-
bration block label.
5. If the value is out of tolerance, refer to troubleshooting section 8.2, or consult fac-
tory service.
7.3 SWITCHING ON
Press and release - key
Temperature
Displayed value
10 °C / 50°F
54,5 ± 0,3 mm
15 °C / 59°F
54,7 ± 0,3 mm
20 °C / 68°F
54,9 ± 0,3 mm
25 °C / 77°F
55,1 ± 0,3 mm
30 °C / 86°F
55,3 ± 0,3 mm