ACCUTEMP SERIES 480°F VACUUM DRYING OVEN USER’S MANUAL
- 14 -
Table 2
(Do NOT modify unless instructed by an AI technician.)
Prompt Name
Description
Range (factory value)
Lc: Pass key
When Lc = 9, controller enters into the menu below
3, 9, 27, 567 (0)
ndA: Temp alarm mode
0: with over-temp alarm only
1: with over-temp alarm and under-temp alarm
0-1 (0)
ndT: Timer mode
0: no timer function
1: with timer function, SV window will display dwelling time after
oven temperature reaches target temperature.
2: with timer function, SV window will always display dwelling
time.
0-2 (1)
Hn: Timer unit
0: displays timer in minutes 1: displays timer in hours
0-1 (0)
SPd: Constant
temperature deviation
When PV > SV – SPd, oven is under constant temperature state
0.1-180 (0.9) under °F mode
0.1-100 (0.5) under °C mode
SPT: Constant temp
state buzzing time
Set buzzing time in seconds when oven enters into constant temp
state. Buzzer will sound constantly if SPT is set to 9999.
0-9999 (0) seconds
EST: Dwelling time up
buzzing time
Set buzzing time in seconds when the dwelling time is up. Buzzer
will sound constantly if EST is set to 9999.
0-9999 (60) seconds
EH: Continue to heat
after dwelling time ends
0: Turns off heating when dwelling time is over 1: Continues to keep
the oven at constant temperature, even after dwelling time is over.
0-1 (0)
ndo: Output mode
0: Output after dwelling time over. 1: Output after over-temp
2: Output when oven enters into constant temperature state.
0-2 (0)
SPL: Min temperature
Set minimum temperature the controller can go to.
-58 to 122 (32) under °F mode
-50 to 50 (0) under °C mode
SPH: Max temperature
Set maximum temperature the controller can go to.
32-752 (572) under °F mode
SPL to 400 (300) under °C mode
Table 3
Prompt Name
Description
Range (factory value)
Lc: Pass key
When Lc = 27, controller enters into the menu below
3, 9, 27, 567 (0)
FC: Temperature
display mode
0: Celsius 1: Fahrenheit
0-1 (0)
Table 4
Prompt Name
Description
Range (factory value)
Lc: Pass key
When Lc = 567, controller enters into the menu below
3, 9, 27, 567 (0)
rST: Factory reset
0: Do not reset to factory settings 1: Reset to factory settings
0-1 (0)
Содержание AccuTemp Series
Страница 7: ...ACCUTEMP SERIES 480 F VACUUM DRYING OVEN USER S MANUAL 6 3 2 cubic foot model ...
Страница 8: ...ACCUTEMP SERIES 480 F VACUUM DRYING OVEN USER S MANUAL 7 7 5a cubic foot model ...
Страница 25: ...ACCUTEMP SERIES 480 F VACUUM DRYING OVEN USER S MANUAL 24 ...
Страница 27: ...ACCUTEMP SERIES 480 F VACUUM DRYING OVEN USER S MANUAL 26 14 SCHEMATICS AND CONTROLLER CONNECTION ...
Страница 28: ...ACCUTEMP SERIES 480 F VACUUM DRYING OVEN USER S MANUAL 27 ...
Страница 29: ...ACCUTEMP SERIES 480 F VACUUM DRYING OVEN USER S MANUAL 28 ...