2.
Installation
2.1
Initial Check
Packaging
The Serinus 60 is transported in packaging which is specifically designed to minimise the effects of
shock and vibration during transportation. Ecotech recommends that the packaging be kept if there
is a likelihood that the instrument is going to be relocated.
Note:
The red plastic caps that seal the pneumatic connections during transport must be removed
prior to operation.
Opening the Instrument
Check the interior of the instrument with the following steps:
1.
Undo the screws located in the rear panel.
2.
Open the chassis lid by releasing the latch (pressing the button) located on the front panel in the
top left-hand corner, slide the lid backwards.
3.
To completely remove, slide the lid backwards until the rollers line up with the gaps in the track
and pull the lid upwards to remove from the instrument (refer to Figure 4).
4.
Check that all pneumatic and electrical connectors are connected. If not, reconnect.
5.
Check for any visible and obvious damage. If damage exists contact your supplier and follow the
instructions in (Claims for Damaged Shipments) and (Shipping Discrepancies) at the front of this
manual.
Figure 4 – Opening the Instrument
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...